Application of novel calibration scheme based on traceable point-like 22Na sources to various types of PET scanners

Authors :
Toru Takeda
Yasuhiro Wada
Kei Kikuchi
Takahiro Yamada
Keiichi Oda
Yasushi Sato
Mikio Matsumoto
Toshiaki Sasaki
Tomoyuki Hasegawa
Hideo Murayama
Kenta Miwa
Kenta Akimoto
Yutaka Abe
Hiroki Miyatake
Kei Wagatsuma
Source :
2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC).
Publication Year :
2012
Publisher :
IEEE, 2012.

Abstract

Purpose: We have been developing a practical and reliable calibration scheme based on the use of traceable pointlike sources. In using 22Na sources, special care should be taken to avoid the effects of 1.275-MeV γ rays accompanying β+ decays. The purpose of this study is to validate this calibration method with traceable point-like 22Na sources on various types of PET scanners. Method: The traceable point-like 22Na sources with a spherical absorber design used in this study were of the same type as those used in a previous study. The tested PET scanners included one clinical whole-body PET scanner, four types of clinical PET/CT scanners from different manufacturers, and one small-animal PET scanner. The ROI (region of interest) diameter dependence of ROI values were represented with a fitting function, which was assumed to consist of a recovery part due to spatial resolution and a quadratic background part originating from the scattered γ rays. Results: The calibration factors determined using the point-like source were consistent with those by the standard cross-calibration method within ±4%, which was comparable to the uncertainty of the standard cross-calibration method. Conclusion: The novel calibration scheme based on the use of traceable 22Na point-like sources was validated for six types of commercial PET scanners.

Details

Database :
OpenAIRE
Journal :
2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC)
Accession number :
edsair.doi...........430b29877c24cef971db66b7a45801b2
Full Text :
https://doi.org/10.1109/nssmic.2012.6551655