Determination of linear alkylbenzenesulfonates in communal wastewater by means of solid phase microextraction coupled with API-MS and HPLC-FLD
- Source :
- Fresenius' Journal of Analytical Chemistry. 365:674-681
- Publication Year :
- 1999
- Publisher :
- Springer Science and Business Media LLC, 1999.
-
Abstract
- The use of Solid Phase Microextraction (SPME) for the qualitative and quantitative determination of Linear Alkylbenzenesulfonates (LAS) in waste water samples was investigated. A Carbowax/Templated Resin (CW/¶TPR) coated fiber was directly immersed into influent and effluent samples of a sewage treatment plant (STP). Extraction conditions such as time, pH, ion strength were investigated. The extracted LAS were desorbed with a solvent in a specially designed SPME-LC interface for analysis with HPLC-FLD and Electrospray Ionization Mass Spectrometry (ESI-MS). The combination of SPME with ESI-MS proved to be an alternative technique for the LAS determination in waste water. Linear ranges of the external calibration were found from 0.5–100 ng/mL, with detection limits of 0.5 ng/mL for each individual LAS homologue. The reproducibility of the method is 15% (relative standard deviation).
Details
- ISSN :
- 14321130 and 09370633
- Volume :
- 365
- Database :
- OpenAIRE
- Journal :
- Fresenius' Journal of Analytical Chemistry
- Accession number :
- edsair.doi...........4980054620b0cfdd66c1d58b7e7abd03
- Full Text :
- https://doi.org/10.1007/s002160051544