Back to Search
Start Over
Sub-keV resolution detection with Cd<formula><roman>1-x</roman></formula>Zn<formula><roman>x</roman></formula>Te detectors
- Source :
- X-Ray Detector Physics and Applications II.
- Publication Year :
- 1993
- Publisher :
- SPIE, 1993.
-
Abstract
- Preliminary investigations were carried out to evaluate the resolution of Cd1-xZnxTe detectors at temperatures achievable with commercially available, low power Peltier refrigerators. Detectors were in the form of cubes, 2 mm on a side. They were tested using a preamplifier with optical feedback. The input FET was cooled along with the detectors. Resolutions of 409 eV and 326 eV were observed at the 5.9 keV line of Fe-55, at temperatures of -10 degree(s)C and -20 degree(s)C, respectively. Results indicate that a straightforward extension of this work will lead to resolutions well below 200 eV.© (1993) COPYRIGHT SPIE--The International Society for Optical Engineering. Downloading of the abstract is permitted for personal use only.
Details
- ISSN :
- 0277786X
- Database :
- OpenAIRE
- Journal :
- X-Ray Detector Physics and Applications II
- Accession number :
- edsair.doi...........4c981637b9a862ccb666ae3fed5314d9
- Full Text :
- https://doi.org/10.1117/12.164731