Relative efficiency calibration of a Si(Li) electron spectrometer

Authors :
L.J. Jardine
C.M. Lederer
Source :
Nuclear Instruments and Methods. 120:515-520
Publication Year :
1974
Publisher :
Elsevier BV, 1974.

Abstract

A method for determination of the relative efficiency of Si(Li) electron spectrometers with the sources 180m Hf, 207 Bi, and 210 At is described. New measurements of some relative γ-ray (photon) intensities in 180m Hf and 207 Bi are given. Electron line intensities for all three sources, calculated from the decay schemes, γ-ray intensities, and theoretical internal conversion coefficients, are tabulated.