Impacto da pureza dos filtros de alumínio no valor de Camada Semi-Redutora em radiologia convencional e mamografia

Authors :
Gabriela Hoff
Nathan Willig Lima
Source :
Brazilian Journal of Radiation Sciences. 4
Publication Year :
2016
Publisher :
Sociedade Brasilieira de Protecao Radiologica - SBPR, 2016.

Abstract

Half Value Layer (HVL) test is an important technical information in conventional radiology once it is the only test which allows us to quantify beam quality in a complete way. This work promotes a preliminary analysis of the impact of aluminum alloy composition in the determination of HVL value taking into account a realistic geometry. Deterministic calculations and Monte Carlo Method were applied in order to make the proposed analysis. Six aluminum alloys (99% pure) commercialized and pure aluminum were used as comparative basis for both methods. Experimental data were collected in conventional radiology, using tungsten target and aluminum filter for two peak tensions (66 kVp, 81 kVp), and in mammography, using three target filter combinations( molybdenum-molybdenum, molybdenum-rhodium, tungsten-rhodium) for three peak tensions (25 kVp, 30 kVp and 35 kVp). Spectra used in the deterministic calculations were taken from Report No 78 of Institute of Physics and Engineering in Medicine, 1997. Results show that Mammography is more sensitive than conventional X-ray equipment to impurities in aluminum alloys (up to 34.65% variation for W-Rh combination with 30 kVp).

Details

ISSN :
23190612
Volume :
4
Database :
OpenAIRE
Journal :
Brazilian Journal of Radiation Sciences
Accession number :
edsair.doi...........5669aef8ee1dd1b3728d2dd4b29d312f
Full Text :
https://doi.org/10.15392/bjrs.v3i2a.92