Back to Search Start Over

High resolution position-sensitive proportional counter

Authors :
R.G.H. Robertson
R.G. Markham
Source :
Nuclear Instruments and Methods. 129:131-140
Publication Year :
1975
Publisher :
Elsevier BV, 1975.

Abstract

A counter giving high resolution for lightly ionizing particles incident at 45° has been designed and placed in routine operation in an Enge split-pole magnetic spectrograph. Use of a lumped-constant delay line and inclined cathode pickup stripes results in a very compact unit with high transmission; the collection and readout geometry results in virtual elimination of straggling (energy-loss fluctuations) as a source of resolution degradation. Spectra with total line widths less than 0.25 mm fwhm have been observed with 35 MeV protons.

Details

ISSN :
0029554X
Volume :
129
Database :
OpenAIRE
Journal :
Nuclear Instruments and Methods
Accession number :
edsair.doi...........568e1ed62145cff29c138dd447076047
Full Text :
https://doi.org/10.1016/0029-554x(75)90122-6