Back to Search Start Over

Search for an optimum time response of spark counters

Authors :
T. Kress
J Eschke
Mihai Petrovici
A. Gobbi
P. Koczon
A. Devismes
N. Herrmann
K. D. Hildenbrand
Ch. Finck
Source :
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 482:179-191
Publication Year :
2002
Publisher :
Elsevier BV, 2002.

Abstract

A spark counter of the type developed by Pestov has been tested with the aim of searching for an optimum time response function, changing voltage, content of noble and quencher gases, pressure and energy-loss. Replacing the usual argon by neon has brought an improvement of the resolution and a significant reduction of tails in the time response function. It has been proven that a counter as long as 90 cm can deliver, using neon gas mixture, a time resolution σ ps with about 1% absolute tail and an efficiency of about 90%.

Details

ISSN :
01689002
Volume :
482
Database :
OpenAIRE
Journal :
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Accession number :
edsair.doi...........69e776064fac9b5c2ca3fc53ccedcc6a
Full Text :
https://doi.org/10.1016/s0168-9002(01)01622-9