Back to Search Start Over

Comparison of independent optical frequency measurements using a portable iodine-stabilized nd:yag laser

Comparison of independent optical frequency measurements using a portable iodine-stabilized nd:yag laser

Authors :
Atsushi Onae
Feng-Lei Hong
Kazuhiko Sugiyama
Jun Ye
Hirokazu Matsumoto
Jun Ishikawa
John L. Hall
Source :
IEEE Transactions on Instrumentation and Measurement. 52:240-244
Publication Year :
2003
Publisher :
Institute of Electrical and Electronics Engineers (IEEE), 2003.

Abstract

A comparison of two independent absolute optical frequency measurements has been carried out between the National Metrology Institute of Japan/National Institute of Advanced Industrial Science and Technology (NMIJ/AIST), Tsukuba, Japan and the JILA (formerly the Joint Institute for Laboratory Astrophysics), Boulder, CO, using a portable iodine-stabilized Nd:YAG laser. The agreement between the two absolute measurements is better than the measurement uncertainty of 6.7/spl times/10/sup -13/ that can be attributed to the reproducibility limitations of the portable laser. This comparison is used to confirm the measured absolute frequency of an iodine-stabilized Nd:YAG laser at NMIJ/AIST (Y3), which is reported to the Consultative Committee for Length (CCL) for the determination of the absolute frequency value of iodine-stabilized Nd:YAG lasers.

Details

ISSN :
00189456
Volume :
52
Database :
OpenAIRE
Journal :
IEEE Transactions on Instrumentation and Measurement
Accession number :
edsair.doi...........6e2b7ca36b84ddc19c8636a4b8342f17
Full Text :
https://doi.org/10.1109/tim.2003.811676