Back to Search Start Over

Development of a phoswich detector for a continuous blood-sampling system

Authors :
Mikio Suga
Hidehiro Iida
Kotaro Minato
Seiichi Yamamoto
K. Tarutani
Hiroshi Watabe
Source :
IEEE Transactions on Nuclear Science. 48:1408-1411
Publication Year :
2001
Publisher :
Institute of Electrical and Electronics Engineers (IEEE), 2001.

Abstract

We have developed a phoswich detector composed of a plastic scintillator and a BGO, for a continuous blood sampling system. The beta particle (positron) from the tube is detected by the plastic scintillator and emits 511 keV gamma photons. The BGO scintillator that is optically coupled to the plastic scintillator detects one of the gamma photons. Since the decay time of the plastic scintillator and the BGO are very different, it is possible to discriminate true (beta+gamma) events from background gamma events. First, the pulse height and pulse shape of plastic scintillator for beta particles and BGO for 511 keV gamma photons were measured and compared to estimate the possibility of the proposed method. Then proposed phoswich detector was fabricated and tested. Absolute sensitivity for Ga-68 (maximum energy of 1.90 MeV) and F-18 (maximum energy of 633 keV) positrons were measured and compared with conventional beta detector in similar size. The absolute sensitivity of the developed detector was O.15 counts/Bq for Ga-68 positrons; at the center of the detector that was approximately 5 times higher sensitivity than the conventional beta detector. The absolute sensitivity of the developed detector for F-18 positrons was 0.017 counts/Bq while that of the conventional beta detector was zero. The count rate of the developed detector was linear up to 10 kcps. The background count rate was small. These results indicate that the developed detector is useful not only for higher energy positrons such as 0-15 but also lower energy positrons such as F-18 or C-11.

Details

ISSN :
15581578 and 00189499
Volume :
48
Database :
OpenAIRE
Journal :
IEEE Transactions on Nuclear Science
Accession number :
edsair.doi...........71a5eaa6f8eb22b0437a2ded073107f0