Back to Search Start Over

X-band Linac for a 6 MeV dual-head radiation therapy gantry

Authors :
Hui Su Kim
SangChul Mun
JongChul Lee
Hyungdal Park
Byung Chul Lee
Seung-Hyun Lee
Byeong-No Lee
Seung-wook Shin
H. S. Song
Jong-Seo Chai
Ki-back Song
Donghyup Ha
Source :
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 852:40-45
Publication Year :
2017
Publisher :
Elsevier BV, 2017.

Abstract

We developed a design for a 6 MeV X-band linear accelerator for radiation therapy in a dual-head gantry layout. The dual-head gantry has two linacs that can be operated independently. Each X-band linac accelerates electron bunches using high-power RF and generates X-rays for radiation therapy. It requires a versatile RF system and pulse sequence to accomplish various radiation therapy procedures. The RF system consists of 9.3 GHz, 2 MW X-band magnetron and associated RF transmission components. A test linac was assembled and operated to characterize its RF performance without beam. This paper presents these results along with a description of the gantry linacs and their operational requirements.

Details

ISSN :
01689002
Volume :
852
Database :
OpenAIRE
Journal :
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Accession number :
edsair.doi...........7391938b5ccf1baed9e210653cf921b1
Full Text :
https://doi.org/10.1016/j.nima.2016.11.034