Low-dose computed tomography of urolithiasis in obese patients: a feasibility study to evaluate image reconstruction algorithms
- Source :
- Diabetes, Metabolic Syndrome and Obesity: Targets and Therapy. 12:439-445
- Publication Year :
- 2019
- Publisher :
- Informa UK Limited, 2019.
-
Abstract
- Purpose: Retrospective evaluation and comparison of image quality generated by low-dose computed tomography (LDCT) from obese patients with urolithiasis using alternative reconstruction algorithms. Materials and methods: Twenty-five obese patients (body mass index [BMI]>25 kg/m2) underwent LDCT scans for suspected urolithiasis. The scans were recompiled using filtered-back projection (FBP), statistical iterative reconstruction (iDose) and iterative model-based reconstruction (IMR). Dose-length product (DLP) and patient details were obtained from the CT dose report and clinical charts, respectively. Objective image noise was assessed by measuring the SD of Hounsfield units (HUs) in defined locations. Additionally, subjective image evaluation was independently performed by two radiologists using a 3-point Likert scale. The inter-reviewer agreement of image quality was calculated. Results: Ureteral concretions were observed in all CT scans, two of which revealed bilateral stones. The assessed patients' mean BMI was 29.29±3.74 kg/m2, and the DLP of the CT scans was 100.04±10.00 mGy*cm. All scans were rated diagnostic with the iDose and iterative model-based reconstructions, whereas 41% of the scans performed with FBP reconstruction were nondiagnostic. With respect to image quality, IMR was superior to iDose and FBP, both in the objective (P
- Subjects :
- Pharmacology
medicine.diagnostic_test
business.industry
Image quality
Low dose
030209 endocrinology & metabolism
Computed tomography
Iterative reconstruction
030204 cardiovascular system & hematology
03 medical and health sciences
chemistry.chemical_compound
0302 clinical medicine
Image reconstruction algorithm
chemistry
Hounsfield scale
Internal Medicine
Image noise
Idose
Medicine
business
Nuclear medicine
Subjects
Details
- ISSN :
- 11787007
- Volume :
- 12
- Database :
- OpenAIRE
- Journal :
- Diabetes, Metabolic Syndrome and Obesity: Targets and Therapy
- Accession number :
- edsair.doi...........752fbe9d1e47544d0c7b8777e6b1fdf6
- Full Text :
- https://doi.org/10.2147/dmso.s198641