Self-Referenced Residual Pressure Measurement Method for Fiber-Optic Pressure Sensor Chip
- Source :
- IEEE Photonics Technology Letters. 26:957-960
- Publication Year :
- 2014
- Publisher :
- Institute of Electrical and Electronics Engineers (IEEE), 2014.
-
Abstract
- We propose a direct and nondestructive self-referenced residual pressure measurement method for MEMS chip of fiber-optic pressure sensor. In the proposed method, two multimode fibers are placed at the center and the edge position of the sealed cavity. With scanning external pressure, two groups of cavity lengths data are obtained by using polarization low-coherence interference demodulator. The intersection point of their linear fitting curves indicates that the diaphragm is in flat condition and residual pressure is determined by the external pressure at this point. We successfully demonstrated the effectiveness of the residual pressure measurement method with measure error 0.049 kPa.
- Subjects :
- Microelectromechanical systems
Multi-mode optical fiber
Optical fiber
Materials science
business.industry
Physics::Medical Physics
Physics::Optics
Chip
Pressure sensor
Atomic and Molecular Physics, and Optics
Electronic, Optical and Magnetic Materials
law.invention
Optics
Pressure measurement
law
Nondestructive testing
Curve fitting
Electrical and Electronic Engineering
business
Subjects
Details
- ISSN :
- 19410174 and 10411135
- Volume :
- 26
- Database :
- OpenAIRE
- Journal :
- IEEE Photonics Technology Letters
- Accession number :
- edsair.doi...........765ae2e06b6730568178901fa74f59c3
- Full Text :
- https://doi.org/10.1109/lpt.2014.2307346