Necessary forward model specification accuracy for basis material decomposition in spectral CT

Authors :
Mats Persson
Hans Bornefalk
Mats Danielsson
Source :
SPIE Proceedings.
Publication Year :
2014
Publisher :
SPIE, 2014.

Abstract

Material basis decomposition in the sinogram domain requires accurate knowledge of the forward model in spectral CT. Misspecifications over a certain limit will result in biased estimates and make quantum limited quantitative CT difficult. We present a method whereby users can determine the degree of allowed misspecification error in a spectral CT forward model, and still have quantification errors that are quantum limited.

Details

ISSN :
0277786X
Database :
OpenAIRE
Journal :
SPIE Proceedings
Accession number :
edsair.doi...........7a1a05e8c948f846ed778ec8002ecafb
Full Text :
https://doi.org/10.1117/12.2043702