Back to Search Start Over

High-performance non-enzymatic glucose sensor based on Ni/Cu/boron-doped diamond electrode

Authors :
Zhenzhi Gong
Can Li
Jingting Luo
Qiuping Wei
Naixiu Hu
Kuangzhi Zheng
Li Ma
Kechao Zhou
Nan Huang
Zhiming Yu
Cheng-Te Lin
Wentao Ye
Source :
Journal of Electroanalytical Chemistry. 841:135-141
Publication Year :
2019
Publisher :
Elsevier BV, 2019.

Abstract

A Ni/Cu/boron-doped diamond (Ni/Cu/BDD) complex electrode for non-enzymatic glucose electrochemical detection was prepared by a simple two-step heat treatment method. Scanning electron microscopy (SEM), Raman spectroscopy and electrochemical workstation were used to characterize the surface morphology, composition and electrochemical properties of the electrode, respectively. The results showed that Ni reacted with BDD under high temperature catalytic conditions forming a porous structure, and stabilized Cu on the surface of BDD due to the superior wettability between Cu and Ni. Compared to Ni/BDD and Cu/BDD electrodes, Ni/Cu/BDD electrode exhibited enhanced catalytic activity in glucose detection, such as an extremely wide detection range (0.022–18.3 mM), high sensitivity (1007.688 μAmM−1 cm−2, which was 1.28 times higher than that of the Ni/BDD electrode), great selectivity and excellent long-term stability (93.3% after one month).

Details

ISSN :
15726657
Volume :
841
Database :
OpenAIRE
Journal :
Journal of Electroanalytical Chemistry
Accession number :
edsair.doi...........7d9a480538b0b14d9ed72adcfd6b9395
Full Text :
https://doi.org/10.1016/j.jelechem.2019.03.043