Back to Search Start Over

Migration of the Duraloc cup at two years

Authors :
B. Frischhut
B. Stöckl
Rainer Biedermann
Martin Krismer
Cornelius Wimmer
M. Sandow
Source :
The Journal of Bone and Joint Surgery. British volume. :51-53
Publication Year :
1999
Publisher :
British Editorial Society of Bone & Joint Surgery, 1999.

Abstract

We carried out 71 primary total hip arthroplasties using porous-coated, hemispherical press-fit Duraloc ‘100 Series’ cups in 68 consecutive patients; 61 were combined with the cementless Spotorno stem and ten with the cemented Lubinus SP II stem. Under-reaming of 2 mm achieved a press-fit. Of the 71 hips, 69 (97.1%) were followed up after a mean of 2.4 years. Migration analysis was performed by the Ein Bild Röntgen Analyse method, with an accuracy of 1 mm. The mean total migration after 24 months was 1.13 mm. Using the definition of loosening as a total migration of 1 mm, it follows that 30 out of 63 cups (48%) were loose at 24 months.

Details

ISSN :
20445377 and 0301620X
Database :
OpenAIRE
Journal :
The Journal of Bone and Joint Surgery. British volume
Accession number :
edsair.doi...........83d46530368a77363dc5933e9721db76