Back to Search Start Over

Chemiluminescence determination of sodium 2-mercaptoethane sulfonate by flow injection analysis using cerium(IV) sensitized by quinine

Authors :
Luis Fermín Capitán-Vallvey
R Acosta Acosta
M. C. Valencia Mirón
Source :
Talanta. 51:1155-1161
Publication Year :
2000
Publisher :
Elsevier BV, 2000.

Abstract

A flow injection method with chemiluminescence (CL) detection is proposed for the determination of sodium 2-mercaptoethane sulfonate (MESNA), a drug often used to reduce the urotoxic effects of antineoplastic alkylating agents. The procedure is based on the reaction of the thiol with Ce(IV) in a sulfuric acid medium and measurement of the CL intensity produced by quinine used as a sensitizer. The optimum conditions for CL emission were investigated. Using the CL peak height as the analytical parameter, MESNA was determined over the concentration range 0.29‐2.21 ng (1.97‐9.85 mg·l 1 ) with a detection limit of 0.21 ng (1.38 mg·l 1 ) and a relative standard deviation (R.S.D.) of 4.1%. The method was applied satisfactorily to the determination of MESNA in pharmaceutical preparations with percentages of recovery between 94 and 105. © 2000 Elsevier Science B.V. All rights reserved.

Details

ISSN :
00399140
Volume :
51
Database :
OpenAIRE
Journal :
Talanta
Accession number :
edsair.doi...........8473677196091ed97032a9c3c360f481
Full Text :
https://doi.org/10.1016/s0039-9140(00)00291-5