The effects of varying the concentrations of cross-linker N, N ′ -methyelene-bis-acrylamide (BIS) from 2% to 4%, and 2-hydroxyethylacrylate (HEA) monomer from 2% to 4% at 5% gelatin on the dose response of BIS–HEA–gelatin (BHEAG) aqueous polymer gel dosimeters have been studied using magnetic resonance imaging (MRI) for relaxation rate ( R 2 ) of water proton. The dosimeters were irradiated with 60Co teletherapy γ -ray source at a constant dose rate, receiving doses up to 30 Gy. The radiation polymerization occurs and increases with increasing initial dose. R 2 is found to decrease mono-exponentially with depth inside the polymer gel and depend strongly upon the initial concentrations of co-monomers (HEA and BIS). Dose–depth map for BHEAG gel was determined for different concentrations of co-monomer (HEA and BIS). The percentage dose depth was also evaluated which leads to a good agreement with the ionization chamber measurements.