Back to Search Start Over

Determination of Pitavastatin in Human Plasma by LC–MS–MS

Authors :
Lihua Wu
Jianzhong Shentu
Jian Liu
Xingjiang Hu
Mingzhu Huang
Xiao Xu
Huili Zhou
Junchun Chen
Source :
Chromatographia. 69:1041-1047
Publication Year :
2009
Publisher :
Springer Science and Business Media LLC, 2009.

Abstract

A simple and rapid LC–MS–MS assay was developed and validated for the quantitative determination of pitavastatin in human plasma. Sample pretreatment involved simple protein precipitation by addition of acetonitrile. Separation was on an Agilent 1.8 μm Zorbax SB-C18 column (150 mm × 4.6 mm) at 25 °C using isocratic elution with methanol–0.1% formic acid in water (85:15, v/v) at a flow rate of 0.4 mL min−1. Detection was performed using electrospray ionization in positive ion multiple reaction monitoring mode by monitoring the ion transitions m/z 422.0 → 290.1 for pitavastatin, and m/z 330.1 → 192.1 for paroxetine (IS). LC–MS–MS was found to improve the quantitation of pitavastatin in plasma and was successfully applied in pharmacokinetic studies.