Back to Search Start Over

Left atrial appendage occlusion using a Watchman device in a transplanted heart with biatrial anastomosis

Authors :
Kousik Krishnan
Clifford J. Kavinsky
Marie-France Poulin
Burhan Mohamedali
Source :
Catheterization and Cardiovascular Interventions. 93:758-760
Publication Year :
2018
Publisher :
Wiley, 2018.

Abstract

A percutaneous left atrial appendage (LAA) occlusion using a Watchman device was performed in a patient with prior biatrial anastomosis orthotopic heart transplant (OHT). Due to the anatomical changes following biatrial anastomosis, transesophageal echocardiography (TEE) imaging and transseptal puncture become technically challenging. We describe how the standard LAA views on TEE were obtained for device sizing, and how the transseptal puncture was modified to cross the scarred septum. This case demonstrates that LAA occlusion using a Watchman device in biatrial anastomosis OHT is safe and can successfully be done.

Details

ISSN :
1522726X and 15221946
Volume :
93
Database :
OpenAIRE
Journal :
Catheterization and Cardiovascular Interventions
Accession number :
edsair.doi...........916238228c75f5bedb0d7a77aac471c2
Full Text :
https://doi.org/10.1002/ccd.28027