Back to Search
Start Over
High sensitive detections of Norovirus DNA and IgG by using multi-SiNW-FET biosensors
- Source :
- 2015 Transducers - 2015 18th International Conference on Solid-State Sensors, Actuators and Microsystems (TRANSDUCERS).
- Publication Year :
- 2015
- Publisher :
- IEEE, 2015.
-
Abstract
- This paper reports ultrasensitive detections of DNA and protein by multi-silicon-nanowire field-effect-transistor (SiNW-FET) biosensors. Single-SiNW-FET, double-SiNW-FET and quad-SiNW-FET have been designed on a single chip and fabricated by a simple top-down fabrication process. Compared with single-SiNW-FET, the multi-SiNW-FET biosensor shows a higher ability of decreasing the noise by averaging the source-drain currents. Both Norovirus DNA and IgG have been detected with the sensors at concentrations as low as 1 fM and 10 fM respectively. The sensor also shown high specificity in 1bp mismatch DNA disturbance.
Details
- Database :
- OpenAIRE
- Journal :
- 2015 Transducers - 2015 18th International Conference on Solid-State Sensors, Actuators and Microsystems (TRANSDUCERS)
- Accession number :
- edsair.doi...........91d19875ad60803e1695abf97ffc63e4
- Full Text :
- https://doi.org/10.1109/transducers.2015.7181230