Wireless esophageal catheter dedicated to respiratory diseases diagnostic

Authors :
François Bellemare
Tommy Desilets
Mohamad Sawan
Source :
ISCAS
Publication Year :
2006
Publisher :
IEEE, 2006.

Abstract

We present the design and test of a portable wireless catheter system combining the simultaneous assessment of the transdiaphragmatic pressure and EMG of the diaphragm. The esophageal catheter includes two micro fabricated pressure sensors and five platinum ring electrodes. The low noise analog front-end features an electrode DC mismatch correction circuitry, a selectable gain (58 dB-97 dB) as well as a high CMRR (80.3 dB). Pressure signals are sampled at 1 kHz while EMG at 4 kHz. The data is transmitted over a wireless Bluetooth/spl reg/ connection towards a computing host to display the received signals in real time. Testing of the device demonstrates its reliability to process pressure and EMG signals and to be used as a stand-alone respiratory parameters assessment device.

Details

Database :
OpenAIRE
Journal :
2006 IEEE International Symposium on Circuits and Systems
Accession number :
edsair.doi...........95ae666521f2246a0361061cb7c73754
Full Text :
https://doi.org/10.1109/iscas.2006.1693151