Development of a high-resolution scintillator-based area detector for neutrons

Authors :
S. Sato
K. Morimoto
Chun-Keung Loong
Jun-ichi Suzuki
Tomohiro Adachi
A. Gorin
S. Suzuki
Fuyuki Tokanai
I. Manuilov
Michihiro Furusaka
K. Sakai
A. Ryazantsev
Hidenori Sagehashi
Naoya Sakamoto
K Ikeda
Takayuki Oku
Hirohiko M. Shimizu
Hiromasa Miyasaka
Kenneth C. Littrell
K. Kuroda
Y. Kiyanagi
Source :
Journal of Applied Crystallography. 36:820-825
Publication Year :
2003
Publisher :
International Union of Crystallography (IUCr), 2003.

Abstract

We applied a scintillator-based area detector to the neutron time-of-flight method by developing a new electronics system for this detector, which detector consists of a thin plate of ZnS(Ag)+6Li scintillator optically coupled to wavelength shifter. The spatial resolution was about 0.4mm which is consistent with a fiber width of 0.4 mm. The detection efficiency for thermal neutrons was about 6%. A good-resolution radiographic-image as a function of flight time was obtained.