Back to Search Start Over

Spectroscopic Measurements of Oxygen Saturation in the Retina

Authors :
Scott A. Mathews
Jessica C. Ramella-Roman
Source :
IEEE Journal of Selected Topics in Quantum Electronics. 13:1697-1703
Publication Year :
2007
Publisher :
Institute of Electrical and Electronics Engineers (IEEE), 2007.

Abstract

We introduce a new type of retinal oximeter able to capture multiple spectroscopic sensitive images of the retina in a single snapshot. The core of our apparatus is a multiaperture camera that can be easily interfaced with a fundus camera system. A lenslet array divides the image collected by a fundus camera into identical subimages. A bandpass filter array in front of the lens system is used to collect the subimages only at the wavelengths of interest. Two different instruments were designed based on this simple principle. The first system can collect six different images, while the second system can collect up to 18 different images in a single snapshot.

Details

ISSN :
1077260X
Volume :
13
Database :
OpenAIRE
Journal :
IEEE Journal of Selected Topics in Quantum Electronics
Accession number :
edsair.doi...........b09fe8aad3ff72ca4cf0661df1438808
Full Text :
https://doi.org/10.1109/jstqe.2007.911312