Back to Search Start Over

Design of a fast neutron coincidence counter

Authors :
N. Ensslin
M.S. Krick
R.C. Byrd
M.C. Miller
W. C. Feldman
P.A. Russo
Thomas H. Prettyman
Source :
Applied Radiation and Isotopes. 48:1549-1555
Publication Year :
1997
Publisher :
Elsevier BV, 1997.

Abstract

Monte Carlo simulations were carried out to simulate the response of a boron-loaded plastic scintillator array. These detectors offer potential advantages compared to moderated 3He systems because of their fast response and high efficiency. Tallies were made of neutron capture probability as a function of time and the results were coupled to assay variance estimates to evaluate detector performance. Orders-of-magnitude reductions in count time relative to a 50% efficient thermal counter are possible, in principle, for high (α,n) samples.

Details

ISSN :
09698043
Volume :
48
Database :
OpenAIRE
Journal :
Applied Radiation and Isotopes
Accession number :
edsair.doi...........b2c729bff90261f95adda2d93f80cda5