Back to Search Start Over

Time stability of liquid xenon photoionization detectors

Authors :
K. Hasuike
Tadayoshi Doke
Jun Kikuchi
T. Takahashi
Hiroyuki Okada
Kimiaki Masuda
M Shinoda
K. Terasawa
Source :
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 374:188-192
Publication Year :
1996
Publisher :
Elsevier BV, 1996.

Abstract

We measured the collected charges, the energy resolution and their time stability using an ionization chamber filled with liquid xenon doped with trimethylamine (TMA) as photoionization molecules. When liquid xenon was doped with a low concentration of TMA purified at −110°C, the chamber was operated stably at −110°C for about 50 h.

Details

ISSN :
01689002
Volume :
374
Database :
OpenAIRE
Journal :
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Accession number :
edsair.doi...........b44cbc0c395c4c6c158920084e79ccef
Full Text :
https://doi.org/10.1016/0168-9002(96)00079-4