Back to Search
Start Over
Determination of Migration of Hazardous Elements from Non-Stick Pan Materials in Five Food Simulants by ICP-MS
- Source :
- Materials Science Forum. 850:72-77
- Publication Year :
- 2016
- Publisher :
- Trans Tech Publications, Ltd., 2016.
-
Abstract
- A method for the determination of seven hazardous elements in Non-stick Pan Materials in five food simulants by inductively coupled plasma mass spectrometry (ICP-MS) was established. Using this method, the experiments were carried out on the samples of imitated migratory elements (Cr、Mn、Ni、Cu、As、Pb、Cd) in food simulants of water, 4%(W/V) acetic acid solution, 10%(V/V) ethanol solution, 2g/L sodium chloride solution and olive oil. The sample preparation method of five food simulants was confirmed through experiment. Experiments by resolution, stability, sensitivity and other parameters were adjusted tuning instrument to determine the instrument to achieve the best working condition. The linear range of this methods was 0.1~50ng/ml. For all elements, the correlation coefficient of the calibration curves was above 0.9990 and the detection limits were within the range of 0.008~0.12ng/ml. Most of the recoveries at 1ng/ml concentration level was from 85.34%~112.3%, while the RSD was from 0.37%~4.1%.
- Subjects :
- Detection limit
Chromatography
Correlation coefficient
Resolution (mass spectrometry)
Calibration curve
Mechanical Engineering
Sodium
Analytical chemistry
chemistry.chemical_element
010501 environmental sciences
Condensed Matter Physics
01 natural sciences
Linear range
chemistry
Mechanics of Materials
General Materials Science
Sample preparation
Inductively coupled plasma mass spectrometry
0105 earth and related environmental sciences