Back to Search Start Over

Quench Propagation Kinetics Within ‘Iseult/INUMAC’ Whole Body 11.7 T MRI Magnet Shielding Coils

Authors :
P. Fazilleau
C Berriaud
F P Juster
Source :
IEEE Transactions on Applied Superconductivity. 20:1818-1822
Publication Year :
2010
Publisher :
Institute of Electrical and Electronics Engineers (IEEE), 2010.

Abstract

In the framework of the French-German consortium Iseult/INUMAC, focused on very-high-magnetic-field molecular imaging, the development of an 11.7 T whole body MRI magnet has been launched in 2006. This magnet is mainly composed of one wet and cryostable main coil and two fully impregnated coils that shield the magnetic field in order to meet the security requirements on the fringe field. To protect the magnet in case of quench, the stored magnetic energy (338.5 MJ) is dissipated in an external resistor and inside the two shielding coils windings. It is thus an important issue to master the normal zone propagation kinetics in those coils. This paper presents the 3d finite-element computations carried out to treat that problem and the more significant obtained results.

Details

ISSN :
15582515 and 10518223
Volume :
20
Database :
OpenAIRE
Journal :
IEEE Transactions on Applied Superconductivity
Accession number :
edsair.doi...........bcabe66f1db239013757954ade9bbfc5
Full Text :
https://doi.org/10.1109/tasc.2010.2041759