Back to Search Start Over

Evaluation of a new, wireless pulse oximetry monitoring system in infants: the BBA bootee

Authors :
Christian Dageville
Christian Terlaud
Dominique Ronayette
Marc Lubrano
Eric Mallet
Larissa I. Netchiporouk
Yves Brusquet
Francis Klefstad-Sillonville
David Blanc
Yves Rimet
Olivier Lerda
Jean-Luc Weber
Jérôme Silve
Caroline Rambaud
Source :
4th International Workshop on Wearable and Implantable Body Sensor Networks (BSN 2007) ISBN: 9783540709930, BSN
Publication Year :
2007
Publisher :
Springer Berlin Heidelberg, 2007.

Abstract

A new system for wireless monitoring of pulse oximetry (SpO2) was developed on the basis of a Nonin OEM III oximetry module and a radiofrequency transceiver. The electronic unit and the power supply was integrated into a specially designed infant shoe named BBA bootee. Clinical evaluation of the system revealed a good agreement between the pulse oximetry data transmitted by the BBA bootee and those recorded simultaneously by a reference monitor. The comparative data collected in 39 babies yielded a mean (bias SD) value of (−1.1 1.9)% for SpO2 and (−2 8) beats per minute for heart rate. Use of an integrated accelerometer/actimeter to reduce the motion artifact is addressed as well as ergonomics of the sensor-supporting garment.

Details

ISBN :
978-3-540-70993-0
ISBNs :
9783540709930
Database :
OpenAIRE
Journal :
4th International Workshop on Wearable and Implantable Body Sensor Networks (BSN 2007) ISBN: 9783540709930, BSN
Accession number :
edsair.doi...........cec6e8da4b9c67a9655ce6f260834912
Full Text :
https://doi.org/10.1007/978-3-540-70994-7_25