Back to Search Start Over

37Cl/35Cl isotope ratio analysis in perchlorate by ion chromatography/multi collector -ICPMS: Analytical performance and implication for biodegradation studies

Authors :
Ludwik Halicz
Faina Gelman
Zeev Ronen
Yevgeni Zakon
Source :
Chemosphere. 184:192-196
Publication Year :
2017
Publisher :
Elsevier BV, 2017.

Abstract

In the present study we propose a new analytical method for 37Cl/35Cl analysis in perchlorate by Ion Chromatography(IC) coupled to Multicollector Inductively Coupled Plasma Mass Spectrometry (MC-ICPMS). The accuracy of the analytical method was validated by analysis of international perchlorate standard materials USGS-37 and USGS -38; analytical precision better than ±0.4‰ was achieved. 37Cl/35Cl isotope ratio analysis in perchlorate during laboratory biodegradation experiment with microbial cultures enriched from the contaminated soil in Israel resulted in isotope enrichment factor e37Cl = −13.3 ± 1‰, which falls in the range reported previously for perchlorate biodegradation by pure microbial cultures. The proposed analytical method may significantly simplify the procedure for isotope analysis of perchlorate which is currently applied in environmental studies.

Details

ISSN :
00456535
Volume :
184
Database :
OpenAIRE
Journal :
Chemosphere
Accession number :
edsair.doi...........cfae1db771e2ba40743a08476256f9d3
Full Text :
https://doi.org/10.1016/j.chemosphere.2017.05.145