Energy-discrimination x-ray computed tomography system utilizing a scanning cadmium-telluride detector

Authors :
Akira Ogawa
Manabu Watanabe
Jun Onagawa
Kiyomi Takahashi
Shigehiro Sato
Eiichi Sato
Keitaro Hitomi
Ablajan Abduraxit
Toshiyuki Enomoto
Source :
SPIE Proceedings.
Publication Year :
2010
Publisher :
SPIE, 2010.

Abstract

An energy-discrimination K-edge x-ray computed tomography (CT) system is useful for controlling the image contrast of a target region by selecting both the photon energy and the energy width. The CT system has an oscillation-type linear cadmium telluride (CdTe) detectror. CT is performed by repeated linear scans and rotations of an object. Penetrating x-ray photons from the object are detected by a CdTe detector, and event signals of x-ray photons are produced using charge-sensitive and shaping amplifiers. Both photon energy and energy width are selected out using a multichannel analyzer, and the number of photons is counted by a counter card. In energy-discrimination CT, the tube voltage and tube current were 80 kV and 20 μA, respectively, and the x-ray intensity was 1.92 μGy/s at a distance of 1.0 m from the source and a tube voltage of 80 kV. The energy-discrimination CT was carried out by selecting x-ray photon energies.