System design and implementation of the detector assembly for the Astro-H soft x-ray spectrometer

Authors :
F. S. Porter
Phillip A. Goodwin
D. McCammon
Meng P. Chiao
S. Shuman
C. W. Hobson
R. L. Kelley
Daniel S. McGuinness
Tomomi Watanabe
Joseph S. Adams
Caroline A. Kilbourne
Samuel J. Moseley
Source :
SPIE Proceedings.
Publication Year :
2016
Publisher :
SPIE, 2016.

Abstract

The soft x-ray spectrometer (SXS) onboard Astro-H presents to the science community unprecedented capability (> 7 eV at 6 keV) for high-resolution spectral measurements in the range of 0.5 – 12 keV to study extended celestial sources. At the heart of this SXS is the x-ray calorimeter spectrometer (XCS) where detectors (calorimeter array and anticoincidence detector) operate at 50 mK, the bias circuit operates at nominal 1.3 K, and the first stage amplifiers operate at 130 K, all within a nominal 20 cm envelope. The design of the detector assembly in this XCS originates from the Astro-E x-ray spectrometer (XRS) and lessons learned from Astro-E and Suzaku. After the production of our engineering model, additional changes were made in order to improve our flight assembly process for better reliability and overall performance. In this poster, we present the final design and implementation of the flight detector assembly, show comparison of parameters and performance to Suzaku’s XRS, and list susceptibilities to other subsystems as well as our lessons learned.

Details

ISSN :
0277786X
Database :
OpenAIRE
Journal :
SPIE Proceedings
Accession number :
edsair.doi...........df98ec69241648e8f87584cdb6f22c70
Full Text :
https://doi.org/10.1117/12.2231897