Back to Search
Start Over
High Resolution Gamma Ray Spectroscopy at MHz Counting Rates With LaBr<formula formulatype='inline'><tex Notation='TeX'>$_{3}$</tex> </formula> Scintillators for Fusion Plasma Applications
- Source :
- IEEE Transactions on Nuclear Science. 60:1408-1415
- Publication Year :
- 2013
- Publisher :
- Institute of Electrical and Electronics Engineers (IEEE), 2013.
-
Abstract
- High resolution γ-ray spectroscopy measurements at MHz counting rates were carried out at nuclear accelerators, combining a LaBr 3(Ce) detector with dedicated hardware and software solutions based on digitization and off-line analysis. Spectra were measured at counting rates up to 4 MHz, with little or no degradation of the energy resolution, adopting a pile up rejection algorithm. The reported results represent a step forward towards the final goal of high resolution γ-ray spectroscopy measurements on a burning plasma device.
Details
- ISSN :
- 15581578 and 00189499
- Volume :
- 60
- Database :
- OpenAIRE
- Journal :
- IEEE Transactions on Nuclear Science
- Accession number :
- edsair.doi...........e216b8311c9165198e11f3bd95db7f5d
- Full Text :
- https://doi.org/10.1109/tns.2013.2252189