Back to Search Start Over

Investigation of the Scanning Angle and the Orientation of the Absorber Effects on the Photoacoustic Tomography Imaging

Authors :
Jian Rong
Li Xin Liu
Bing Zhang Chen
Xiao Chun Zhong
Lin Huang
Kui Lu
Source :
Advanced Materials Research. :1373-1377
Publication Year :
2012
Publisher :
Trans Tech Publications, Ltd., 2012.

Abstract

In this paper, based on circular scanning configuration of photoacoustic tomography (PAT), we have tested the effects of the stepping angle of scanning and the relationship between the scanning range and the orientation of the absorber on the quality of the reconstructed image. The experiment results have shown that the reconstructed image would be influenced more by artifacts as the increase of the stepping angle of scanning, and for the limited-view scanning, while the relationship of the scanning path and the orientation of the absorber is parallel, compared with vertical, we can use the data got from less detecting position to reconstruct image clearly.

Details

ISSN :
16628985
Database :
OpenAIRE
Journal :
Advanced Materials Research
Accession number :
edsair.doi...........ea24985d260e97734be7a00b62ae204d
Full Text :
https://doi.org/10.4028/www.scientific.net/amr.463-464.1373