Using a pulsed laser beam to investigate the feasibility of sub-pixel position resolution with time-correlated transient signals in 3D pixelated CdZnTe detectors
- Source :
- Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 867:7-14
- Publication Year :
- 2017
- Publisher :
- Elsevier BV, 2017.
-
Abstract
- We evaluated the X-Y position resolution achievable in 3D pixelated detectors by processing the signal waveforms readout from neighboring pixels. In these measurements we used a focused light beam, down to 10 μ m, generated by a ∼1 mW pulsed laser (650 nm) to carry out raster scans over selected 3 × 3 pixel areas, while recording the charge signals from the 9 pixels and the cathode using two synchronized digital oscilloscopes.