An impedance method for the measurement of liquid hold-up in two-phase flow

Authors :
Paolo Andreussi
M. Messia
A. Di Donfrancesco
Source :
International Journal of Multiphase Flow. 14:777-785
Publication Year :
1988
Publisher :
Elsevier BV, 1988.

Abstract

A liquid hold-up gauge based on the measurement of the electrical impedance has been developed for application in gas-liquid pipe flow. The gauge consists of two ring electrodes mounted flush to the pipe wall. The impedance (capacitance or conductance) seen by the electrodes depends on the distance between them and on the liquid hold-up. For distances above three tube diameters, the impedance is independent of the flow configuration for all separated flow patterns and, with good approximation, also for intermittent flows. Moreover, capacitance or conductance are linearly related to the liquid hold-up. The impedance under bubble flow conditions closely follows the theoretical predictions due to Maxwell. Also for the other flow configurations (annular, stratified, intermittent) the results of static and/or dynamic calibration agree closely with theoretical models.