Back to Search Start Over

X-Band Elektronenspinresonanz mit Blochscher meßanordnung

Authors :
P. Dullenkopf
A. Witte
G. Laukien
Source :
Applied Physics. 2:63-70
Publication Year :
1973
Publisher :
Springer Science and Business Media LLC, 1973.

Abstract

The use of Bloch-type arrangements (induction principle) in ESR spectrometry is superior over conventional reflection arrangements. The advantage results from the better decoupling between transmitter and receiver, resulting in a higher stability of the overall system. A cylindrical cavity sustaining the dual degenerate TE 112 0 modes showed best performance. With such cavities, decoupling-values of more than—110 dB are possible due to special devices. The spectrometer has an overall noise-figure of 8.3 dB, and 6 dB with an additional parametric amplifier. The decoupling device allows to couple circularly polarized waves into the cavity. Thus the sign of the spectroscopic splitting constant may be determined.

Details

ISSN :
14320630 and 03403793
Volume :
2
Database :
OpenAIRE
Journal :
Applied Physics
Accession number :
edsair.doi...........f988562057dea1df084fad4873bee30e
Full Text :
https://doi.org/10.1007/bf00934175