Back to Search Start Over

Respiratory gating algorithm helps to reconstruct more accurate electroanatomical maps during atrial fibrillation ablation performed under spontaneous respiration

Authors :
Béla Merkely
N Parazs
Attila Tóth
Júlia Karády
Gábor Széplaki
István Osztheimer
László Gellér
Tamás Tahin
Szabolcs Szilágyi
Orsolya Mária Kovács
Emin Evren Özcan
Pál Maurovich-Horvat
Source :
Journal of Interventional Cardiac Electrophysiology. 46:153-159
Publication Year :
2016
Publisher :
Springer Science and Business Media LLC, 2016.

Abstract

Electroanatomical mapping is a useful tool during the ablation of atrial fibrillation. Respiratory movement might influence the mapping accuracy and merging. This study aims to investigate the effect of respiratory gating on the accuracy of magnetic-field-based electroanatomical mapping under spontaneous respiration. Fifty-one consecutive patients (35 male, aged 30–78 years) who underwent left atrial radiofrequency catheter ablation due to atrial fibrillation were included. Electroanatomical mapping was performed with CARTO 3 System under conscious sedation. Respiratory gating was achieved with the AccuResp algorithm (Biosense Webster). Average surface match and maximum distance of the pre-acquired and electroanatomical maps, as well as left atrial volume, were recorded with and without respiratory gating after merging. The average surface match of the electroanatomical map with the left atrial reconstruction was significantly better with respiratory gating than without using the algorithm (3.81 ± 1.09 vs 4.11 ± 1.61 mm, p = 0.0119). It was not dependent of the rhythm during mapping or the image modality used for left atrial reconstruction. The maximal distance between the two maps did not depend on the use of the algorithm (19.81 ± 6.24 mm for gated and 20.87 ± 7.99 mm for non-gated, p = 0.3161). Left atrial volume of the map was significantly lower when using the respiratory compensation module (106.3 ± 31.6 vs 127.0 ± 36.4 ml, p

Details

ISSN :
15728595 and 1383875X
Volume :
46
Database :
OpenAIRE
Journal :
Journal of Interventional Cardiac Electrophysiology
Accession number :
edsair.doi.dedup.....02d26446621c731e57ecc427799555a6
Full Text :
https://doi.org/10.1007/s10840-016-0105-x