Double sensing by atrial automatic tachycardia-terminating pulse generator
- Source :
- Pacing and clinical electrophysiology : PACE. 10(1 Pt 1)
- Publication Year :
- 1987
-
Abstract
- This report describes two cases of double sensing of the P and QRS signals by an implanted atrial bipolar automatic burst tachycardia-terminating puise generator (Inlermedics CyherTach 60) used for the treatment of supraventricular tachycardia. Double sensing occurred during normai sinus rhythm at a rate slower than the tachycardia detection criterion and caused inappropriate delivery of burst stimulation. These observations underscore the importance of the far-field QRS signal in the detection of supraventricular tachycardia by automatic puise generators sensing the atrial eiectrogram.