Back to Search
Start Over
Using additional pressure control lines when connecting a continuous renal replacement therapy device to an extracorporeal membrane oxygenation circuit
- Source :
- BMC Nephrology, BMC Nephrology, Vol 19, Iss 1, Pp 1-10 (2018)
- Publication Year :
- 2018
- Publisher :
- BioMed Central, 2018.
-
Abstract
- Background The introduction of a continuous renal replacement therapy (CRRT) device into the extracorporeal membrane oxygenation (ECMO) circuit is widely used. However, excessive pressure transmitted to the CRRT device is a major disadvantage. We investigated the effects of using additional pressure control lines on the pressure and the lifespan of the CRRT circuit connected to the ECMO. Methods This is an observational study using prospectively collected data from consecutive patients receiving CRRT connected into the ECMO circuit at a university-affiliated, tertiary hospital from January 2013 to December 2016. The CRRT circuit was connected into the ECMO circuit through the Luer Lock connection without an additional pressure control line in 16 patients (9%, no line group), an additional pressure control line on the inlet line in 36 patients (23%, single line group), and additional pressure control lines on both the inlet and outlet lines in 118 patients (77%, double line group). The outcome measures of interest were compared among the three groups. Results The median access pressure was higher in the no line group compared to the groups. However, median filter pressure, effluent pressure, and return pressure were higher in the double line group compared to the other groups. There were no significant differences in platelets, lactate dehydrogenase, and plasma hemoglobin among the 3 groups over the time period studied. Median lifespan of the CRRT circuits in the double line group was 45.0 (29.0–63.7) hours, which was higher compared to 21.8 (11.6–31.8) hours in the no line group and 23.0 (15.0–34.6) hours in the single line group, respectively. In addition, in-hospital mortality was lower in the double line group (48.3%) compared to the no line group (68.8%) and the single line group (75.0%). Conclusions Additional tubing can be considered a simple and safe method for pressure control and lengthening circuit survival when connecting the CRRT device to the ECMO circuit.
- Subjects :
- Male
medicine.medical_specialty
Time Factors
Continuous Renal Replacement Therapy
medicine.medical_treatment
Line group
030232 urology & nephrology
030204 cardiovascular system & hematology
lcsh:RC870-923
03 medical and health sciences
0302 clinical medicine
Extracorporeal Membrane Oxygenation
Renal Dialysis
Internal medicine
Extracorporeal membrane oxygenation
medicine
Pressure
Humans
Renal replacement therapy
Hospital Mortality
Prospective Studies
Aged
Retrospective Studies
Pressure control
business.industry
Outcome measures
Shock
Middle Aged
Single line
lcsh:Diseases of the genitourinary system. Urology
Extracorporeal membrane oxygenation (ECMO)
Line (electrical engineering)
Acute kidney injury
Renal Replacement Therapy
Critical care
surgical procedures, operative
Nephrology
Anesthesia
Commentary
Extracorporeal membrane oxygenation circuit
Equipment Failure
Female
business
Respiratory Insufficiency
Dialysis
Research Article
Subjects
Details
- Language :
- English
- ISSN :
- 14712369
- Volume :
- 19
- Database :
- OpenAIRE
- Journal :
- BMC Nephrology
- Accession number :