Bipolar radiofrequency ablation with 2 × 2 electrodes as a building block for matrix radiofrequency ablation:Ex vivoliver experiments and finite element method modelling

Authors :
Yicheng Ni
Guy Marchal
Yansheng Jiang
Jacques Jamart
Yuanbo Feng
Chong Wang
Luc Michel
Stefaan Mulier
Source :
International Journal of Hyperthermia. 31:649-665
Publication Year :
2015
Publisher :
Informa UK Limited, 2015.

Abstract

Size and geometry of the ablation zone obtained by currently available radiofrequency (RF) electrodes is highly variable. Reliability might be improved by matrix radiofrequency ablation (MRFA), in which the whole tumour volume is contained within a cage of x × y parallel electrodes. The aim of this study was to optimise the smallest building block for matrix radiofrequency ablation: a recently developed bipolar 2 × 2 electrode system.In ex vivo bovine liver, the parameters of the experimental set-up were changed one by one. In a second step, a finite element method (FEM) modelling of the experiment was performed to better understand the experimental findings.The optimal power to obtain complete ablation in the shortest time was 50-60 W. Performing an ablation until impedance rise was superior to ablation for a fixed duration. Increasing electrode diameter improved completeness of ablation due to lower temperature along the electrodes. A chessboard pattern of electrode polarity was inferior to a row pattern due to an electric field void in between the electrodes. Variability of ablation size was limited. The FEM correctly simulated and explained the findings in ex vivo liver.These experiments and FEM modelling allowed a better insight in the factors influencing the ablation zone in a bipolar 2 × 2 electrode RF system. With optimal parameters, complete ablation was obtained quickly and with limited variability. This knowledge will be useful to build a larger system with x × y electrodes for MRFA.

Details

ISSN :
14645157 and 02656736
Volume :
31
Database :
OpenAIRE
Journal :
International Journal of Hyperthermia
Accession number :
edsair.doi.dedup.....0fb5fda5fed400889c9b2d1472a50e39
Full Text :
https://doi.org/10.3109/02656736.2015.1046194