Personal Dosimetry in Direct Pulsed Photon Fields with the Dosepix Detector

Authors :
Dennis Haag
Sebastian Schmidt
Patrick Hufschmidt
Gisela Anton
Rafael Ballabriga
Rolf Behrens
Michael Campbell
Franziska Eberle
Christian Fuhg
Oliver Hupe
Xavier Llopart
Jurgen Roth
Lukas Tlustos
Winnie Wong
Hayo Zutz
Thilo Michel
Publication Year :
2021

Abstract

First investigations regarding dosimetric properties of the hybrid, pixelated, photon-counting Dosepix detector in the direct beam of a pulsed photon field (RQR8) for the personal dose equivalent $H\mathrm{_p(10)}$ are presented. The influence quantities such as pulse duration and dose rate were varied, and their responses were compared to the legal limits provided in PTB-A 23.2. The variation of pulse duration at a nearly constant dose rate of about 3.7$\,$Sv/h shows a flat response around 1.0 from 3.6$\,$s down to 2$\,$ms. A response close to 1.0 is achieved for dose rates from 0.07$\,$Sv/h to 35$\,$Sv/h for both pixel sizes. Above this dose rate, the large pixels (220$\,\mu$m edge length) are below the lower limit. The small pixels (55$\,\mu$m edge length) stay within limits up to 704$\,$Sv/h. The count rate linearity is compared to previous results, confirming the saturating count rate for high dose rates.<br />Comment: 6 pages, 4 figures. This work was accepted for publication in IEEE Transactions on Nuclear Science