Back to Search Start Over

Transient out‐of‐range impedance in 'hybrid' implantable cardioverter‐defibrilator system: A case series

Authors :
Wojciech Kwaśniewski
Michał Orszulak
Artur Filipecki
Katarzyna Mizia-Stec
Dagmara Urbańczyk-Swić
Witold Orszulak
Source :
Pacing and Clinical Electrophysiology. 45:270-273
Publication Year :
2021
Publisher :
Wiley, 2021.

Abstract

A retrospective analysis of 60 patients with hybrid ICD systems: Boston Scientific device paired with non-Boston leads. In 10 (17%) patients transient, out-of-range peaks of ventricular pace impedance trend were observed. Probable cause is header-lead interaction incompatibility. This matter is known mainly for pacemakers systems but not for ICDs. Investigation this issue is crucial because consequences in ICD systems are unpredictable and risk might be higher than in pacing systems. This article is protected by copyright. All rights reserved.

Details

ISSN :
15408159 and 01478389
Volume :
45
Database :
OpenAIRE
Journal :
Pacing and Clinical Electrophysiology
Accession number :
edsair.doi.dedup.....14ef590eff428696af1bb58f692afd3d
Full Text :
https://doi.org/10.1111/pace.14401