Transient out‐of‐range impedance in 'hybrid' implantable cardioverter‐defibrilator system: A case series
- Source :
- Pacing and Clinical Electrophysiology. 45:270-273
- Publication Year :
- 2021
- Publisher :
- Wiley, 2021.
-
Abstract
- A retrospective analysis of 60 patients with hybrid ICD systems: Boston Scientific device paired with non-Boston leads. In 10 (17%) patients transient, out-of-range peaks of ventricular pace impedance trend were observed. Probable cause is header-lead interaction incompatibility. This matter is known mainly for pacemakers systems but not for ICDs. Investigation this issue is crucial because consequences in ICD systems are unpredictable and risk might be higher than in pacing systems. This article is protected by copyright. All rights reserved.
- Subjects :
- Male
medicine.medical_specialty
business.industry
General Medicine
Prosthesis Design
System a
Defibrillators, Implantable
Diagnosis, Differential
Equipment Failure Analysis
Internal medicine
Electric Impedance
medicine
Cardiology
Retrospective analysis
Humans
Female
Transient (oscillation)
Cardiology and Cardiovascular Medicine
business
Aged
Retrospective Studies
Subjects
Details
- ISSN :
- 15408159 and 01478389
- Volume :
- 45
- Database :
- OpenAIRE
- Journal :
- Pacing and Clinical Electrophysiology
- Accession number :
- edsair.doi.dedup.....14ef590eff428696af1bb58f692afd3d
- Full Text :
- https://doi.org/10.1111/pace.14401