A cylindrically symmetric magnetic shield for a large-bore 3.0 Tesla magnet
- Source :
- Magnetic resonance in medicine. 29(3)
- Publication Year :
- 1993
-
Abstract
- A 3.0 Tesla, 0.80-m bore magnet replaced our previous 1.9 Tesla, 0.76-m magnet. The 3.0 Tesla replacement magnet had a dipole moment of 1.7 with respect to that of the 1.9 Tesla magnet. The pre-existing cylindrically symmetric passive steel shielding was modified to confine the fringe fields of the replacement magnet to match those of the previous magnet. A cylindrically symmetric inner shield insert of about 20,000 kg was designed, fabricated, and installed. Upon energization of the magnet, the combined shielding met all design criteria. Alternative designs, optimization of cylindrically symmetric designs, and costs of fabrication, are presented herein.
- Subjects :
- Fabrication
Physics::Instrumentation and Detectors
Surface Properties
Instrumentation
Physics::Medical Physics
Astrophysics::Cosmology and Extragalactic Astrophysics
Nuclear magnetic resonance
Optics
Electromagnetic Fields
Radiation Protection
Dipole magnet
Shield
Radiology, Nuclear Medicine and imaging
Physics
Electropermanent magnet
business.industry
Equipment Design
Models, Theoretical
Magnetic Resonance Imaging
Carbon
Dipole
Steel
Magnet
Electromagnetic shielding
Costs and Cost Analysis
Nuclear Medicine Department, Hospital
business
Subjects
Details
- ISSN :
- 07403194
- Volume :
- 29
- Issue :
- 3
- Database :
- OpenAIRE
- Journal :
- Magnetic resonance in medicine
- Accession number :
- edsair.doi.dedup.....184ee80d6e1a164159ebdd9b45e65952