Calibration procedures of the Tore-Supra infrared endoscopes
- Source :
- Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2018, 879, pp.121-133. ⟨10.1016/j.nima.2017.09.035⟩, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Elsevier, 2018, 879, pp.121-133. ⟨10.1016/j.nima.2017.09.035⟩
- Publication Year :
- 2018
- Publisher :
- Elsevier BV, 2018.
-
Abstract
- Five endoscopes equipped with infrared cameras working in the medium infrared range (3–5 μ m) are installed on the controlled thermonuclear fusion research device Tore-Supra. These endoscopes aim at monitoring the plasma facing components surface temperature to prevent their overheating. Signals delivered by infrared cameras through endoscopes are analysed and used on the one hand through a real time feedback control loop acting on the heating systems of the plasma to decrease plasma facing components surface temperatures when necessary, on the other hand for physics studies such as determination of the incoming heat flux . To ensure these two roles a very accurate knowledge of the absolute surface temperatures is mandatory. Consequently the infrared endoscopes must be calibrated through a very careful procedure. This means determining their transmission coefficients which is a delicate operation. Methods to calibrate infrared endoscopes during the shutdown period of the Tore-Supra machine will be presented. As they do not allow determining the possible transmittances evolution during operation an in-situ method is presented. It permits the validation of the calibration performed in laboratory as well as the monitoring of their evolution during machine operation. This is possible by the use of the endoscope shutter and a dedicated plasma scenario developed to heat it. Possible improvements of this method are briefly evoked.
- Subjects :
- Physics
Nuclear and High Energy Physics
Thermonuclear fusion
business.industry
Infrared
Plasma
Tore Supra
01 natural sciences
7. Clean energy
010305 fluids & plasmas
[SPI]Engineering Sciences [physics]
Optics
Heat flux
Shutter
0103 physical sciences
[SPI.OPTI]Engineering Sciences [physics]/Optics / Photonic
Calibration
010306 general physics
business
Instrumentation
ComputingMilieux_MISCELLANEOUS
Overheating (electricity)
Subjects
Details
- ISSN :
- 01689002 and 18729576
- Volume :
- 879
- Database :
- OpenAIRE
- Journal :
- Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
- Accession number :
- edsair.doi.dedup.....1b52affca058e915f0e21cb7df676fd0
- Full Text :
- https://doi.org/10.1016/j.nima.2017.09.035