The treatment planning of segmental, conformal stereotactic radiosurgery utilizing a standard multileaf collimator

Authors :
James M. Balter
Howard M. Sandler
Paul G. Archer
James A. Hayman
Donald A. Ross
Source :
Medical dosimetry : official journal of the American Association of Medical Dosimetrists. 24(1)
Publication Year :
1999

Abstract

Over a period of approximately 3 years, our institution has implemented and refined a system of Stereotactic Radiosurgery (SRS) which utilizes the standard multi leaf collimator (MLC) of the Scanditronix MM50 Racetrack Microtron and treats in an arrangement of segmental “pseudo-arcs.” This system employs a commercial BRW-based stereotactic frame which is mounted to the treatment table. With the exception of the table-mounted frame hardware there have been no modifications to the treatment machine to accommodate these treatments. By use of standard evaluation parameters (e.g., treatment time, planning time, dose conformance and dose heterogeneity ratios) this system compares quite favorably with reported data from institutions treating SRS with either a GammaKnife or a standard linear accelerator with tertiary collimators.