Evaluation of a Newly Developed Rapid Automated Fluorescent Lateral Flow Immunoassay to Detect IgG and IgM Antibodies to Chikungunya Virus

Authors :
Hyeyoung Lee
Sung-Yeon Cho
Joo Hee Jang
Eun Jee Oh
Chulmin Park
Ji Hyeong Ryu
Dong-Gun Lee
Sojeong Yun
Ae Ran Choi
Source :
Infection & Chemotherapy
Publication Year :
2020

Abstract

The ichroma Chikungunya virus (CHIKV) IgG/IgM (Boditech Med Inc., Chuncheon, Korea) is a newly developed rapid lateral flow immunoassay for detection of anti- CHIKV-IgG/IgM. This study conducted with thirty-six anti-CHIKV IgG positive sera, 57 anti-CHIKV IgM positive sera and 163 anti-CHIKV IgG/IgM negative sera which were confirmed by commercial enzyme-linked immunosorbent assays (ELISAs) (Inbios CHIKjj Detectâ„¢ IgM Capture ELISA, Inbios CHIKjj Detectâ„¢ IgG ELISA (InBios International Inc., Seattle, WA, USA), Anti-CHIKV ELISA (IgM), Anti- CHIKV ELISA (IgG) (Euroimmun, Lubeck, Germany)). The ichroma detected all 36 anti-CHIKV IgG and 57 anti-CHIKV IgM positivity (100% sensitivity). For 163 anti-CHIKV IgG/IgM negative sera, the ichroma showed one false positive for IgM (99.4% specificity). The ichroma showed no cross-reactivity and no interference. The ichroma demonstrated good diagnostic performance compared to the current ELISAs.

Details

ISSN :
20932340
Volume :
52
Issue :
4
Database :
OpenAIRE
Journal :
Infectionchemotherapy
Accession number :