Back to Search Start Over

Proposal of a Novel Index for Selection of Optimal Annuloplasty Ring Size for Tricuspid Annuloplication

Authors :
Tomohiro Mizuno
Hirokuni Arai
Naoto Miyagi
Tsuyoshi Hachimaru
Shuhei Fujita
Makoto Tomita
Susumu Manabe
Satoru Kawaguchi
Source :
Circulation Journal. 77:2505-2513
Publication Year :
2013
Publisher :
Japanese Circulation Society, 2013.

Abstract

BACKGROUND Optimal ring size in tricuspid annuloplasty (TAP) surgery to treat functional tricuspid regurgitation (TR) was investigated because optimal ring size remains undefined. METHODS AND RESULTS Sixty seven patients who underwent TAP at our institution were retrospectively studied. Tricuspid Annuloplasty Ring size Index (TARI) was defined as implanted tricuspid annuloplasty ring size divided by body surface area (BSA). Different TARI cut-off values were tested to determine which value produced the greatest difference in TR improvement (TRI=preoperative minus postoperative TR grade) between patients with TARI smaller (group S) and larger (group L) than the cut-off. Group S was also subdivided by ring type: Cosgrove rings (SC) and MC3 rings (SM). TARI and TRI were negatively correlated (r=-0.307). A TARI threshold of 18.9 mm/m(2) produced the greatest and most significant difference (P