Comparison of lead failure manifestation of Biotronik Linox with St. Jude Medical Riata and Medtronic Sprint Fidelis lead

Authors :
Eleni Goulouti
Hildegard Tanner
Stefan Buehler
Laurent Roten
Helge Servatius
Anna Lam
Fabian Noti
Samuel H. Baldinger
Romy Sweda
Jens Seiler
Andreas Haeberlin
Argelia Medeiros-Domingo
Source :
Lam, Anna; Bühler, Stefan Andreas; Goulouti, Eleni; Sweda, Romy; Haeberlin, Andreas; Medeiros Domingo, Argelia; Servatius, Helge Simon; Seiler, Jens; Baldinger, Samuel Hannes; Noti, Fabian; Tanner, Hildegard; Roten, Laurent (2019). Comparison of lead failure manifestation of Biotronik Linox with St. Jude Medical Riata and Medtronic Sprint Fidelis lead. Journal of interventional cardiac electrophysiology, 54(2), pp. 161-170. Springer 10.1007/s10840-018-0486-0
Publication Year :
2019
Publisher :
Springer, 2019.

Abstract

PURPOSE To compare lead failure manifestation and lead performance of the Biotronik Linox/Sorin Vigila defibrillator lead (Linox group) with the St. Jude Medical Riata/Riata ST (Riata group) and Medtronic Sprint Fidelis defibrillator leads (Fidelis group). METHODS We assessed the performance of all aforementioned leads implanted at our center and investigated the manifestation of lead failures. RESULTS Of 93 Linox, 86 Riata, and 81 Fidelis leads implanted at our center, 11 (12%), 22 (26%), and 25 (31%) leads failed during a median follow-up of 46, 61, and 84 months, respectively. Inappropriate shocks were delivered in 64% (Linox), 5% (Riata), and 32% (Fidelis) of lead failures; a device alert was noted in none (Linox), 5% (Riata), and 52% (Fidelis); and lead failure was a coincidental finding in 36% (Linox), 91% (Riata), and 16% (Fidelis) of cases (p

Details

Language :
English
Database :
OpenAIRE
Journal :
Lam, Anna; B&#252;hler, Stefan Andreas; Goulouti, Eleni; Sweda, Romy; Haeberlin, Andreas; Medeiros Domingo, Argelia; Servatius, Helge Simon; Seiler, Jens; Baldinger, Samuel Hannes; Noti, Fabian; Tanner, Hildegard; Roten, Laurent (2019). Comparison of lead failure manifestation of Biotronik Linox with St. Jude Medical Riata and Medtronic Sprint Fidelis lead. Journal of interventional cardiac electrophysiology, 54(2), pp. 161-170. Springer 10.1007/s10840-018-0486-0 <http://dx.doi.org/10.1007/s10840-018-0486-0>
Accession number :
edsair.doi.dedup.....254a08b1aeaee0f61a8c95d3a92b4497
Full Text :
https://doi.org/10.7892/boris.126054