Cytotoxicity and estrogenicity of a novel 3-dimensional printed orthodontic aligner

Authors :
Harris Pratsinis
Spyridon N. Papageorgiou
Nearchos Panayi
Anna Iliadi
Theodore Eliades
Dimitris Kletsas
University of Zurich
Eliades, Theodore
Publication Year :
2022
Publisher :
Elsevier, 2022.

Abstract

Orthodontic aligners printed with in-office 3-dimensional (3D) procedures have been described, but no data on their biocompatibility exist. This study investigates the cytotoxicity and estrogenicity of a 3D-printed orthodontic aligner by assessing its biological and behavioral effects.Ten sets of 1 type of aligner were immersed in sterile deionized water for 14 days, and the cytotoxicity and estrogenicity of released factors were assessed via MTT (3-[4,5-dimethylthiazol-2-yl]-2,5-diphenyltetrazolium bromide) assays on human gingival fibroblasts and the estrogen-sensitive MCF-7 and the estrogen-insensitive MDA-MB-231 breast cancer cell lines. 17β-Estradiol and bisphenol-A were used as positive controls. The statistical analysis of data was performed with generalized linear models at a 0.05 level of significance.No signs of cytotoxicity were seen for the aligner samples for concentrations (v/v) of 20% (P = 0.32), 10% (P = 0.79), or 5% (P = 0.76). The antioxidant activity expressed as the capacity to reduce intracellular levels of reactive oxygen species was not affected in the aligner samples (P = 0.08). No significant estrogenicity was induced by the aligner samples compared with eluents from the negative control for both MCF-7 (P = 0.65) and MDA-MB-231 (P = 0.78). As expected, 17β-Estradiol and bisphenol-A stimulated MCF-7 cell proliferation, whereas no effect was observed on MDA-MB-231 cells.In conclusion, if any factors were released during the 14-day aging of 3D-printed aligners in water, these were not found to be cytotoxic for human gingival fibroblasts and did not affect their intracellular reactive oxygen species levels. Moreover, no estrogenic effects of these putative eluates were observed based on an E-screen assay.

Details

Database :
OpenAIRE
Accession number :
edsair.doi.dedup.....28af97fcca98e469574f5785e01d2f9c
Full Text :
https://doi.org/10.5167/uzh-229695