Back to Search
Start Over
Validation of a Novel Compact System for the Measurement of Lung Volumes
- Source :
- Chest. 159:2356-2365
- Publication Year :
- 2021
- Publisher :
- Elsevier BV, 2021.
-
Abstract
- Background Current techniques for measuring absolute lung volumes rely on bulky and expensive equipment and are complicated to use for the operator and the patient. A novel method for measurement of absolute lung volumes, the MiniBox method, is presented. Research Question Across a population of patients and healthy participants, do values for total lung capacity (TLC) determined by the novel compact device (MiniBox, PulmOne Advanced Medical Devices, Ltd.) compare favorably with measurements determined by traditional whole body plethysmography? Study Design and Methods A total of 266 participants (130 men) and respiratory patients were recruited from five global centers (three in Europe and two in the United States). The study population comprised individuals with obstructive (n = 197) and restrictive (n = 33) disorders as well as healthy participants (n = 36). TLC measured by conventional plethysmography (TLCPleth) was compared with TLC measured by the MiniBox (TLCMB). Results TLC values ranged between 2.7 and 10.9 L. The normalized root mean square difference (NSD) between TLCPleth and TLCMB was 7.0% in healthy participants. In obstructed patients, the NSD was 7.9% in mild obstruction and 9.1% in severe obstruction. In restricted patients, the NSD was 7.8% in mild restriction and 13.9% in moderate and severe restriction. No significant differences were found between TLC values obtained by the two measurement techniques. Also no significant differences were found in results obtained among the five centers. Interpretation TLC as measured by the novel MiniBox system is not significantly different from TLC measured by conventional whole body plethysmography, thus validating the MiniBox method as a reliable method to measure absolute lung volumes.
- Subjects :
- Adult
Male
Pulmonary and Respiratory Medicine
Population
Critical Care and Intensive Care Medicine
Pulmonary function testing
03 medical and health sciences
0302 clinical medicine
Functional residual capacity
Humans
Plethysmograph
Medicine
Lung volumes
030212 general & internal medicine
education
Aged
Whole Body Plethysmography
education.field_of_study
business.industry
Total Lung Capacity
Middle Aged
respiratory system
Healthy Volunteers
United States
Europe
Plethysmography
030228 respiratory system
Background current
Population study
Female
Lung Volume Measurements
Cardiology and Cardiovascular Medicine
business
Nuclear medicine
Subjects
Details
- ISSN :
- 00123692
- Volume :
- 159
- Database :
- OpenAIRE
- Journal :
- Chest
- Accession number :
- edsair.doi.dedup.....28b05d4afdffc904957a4b6a4caf627f
- Full Text :
- https://doi.org/10.1016/j.chest.2021.01.052